EFFECT OF THE GAIT SPEED ON A NEW PROSTHETIC SHANK BELOW KNEE

المؤلفون

  • Zainab H. Zaier Material Engineering Department, College of Engineering, Mustansiriyah University, Baghdad, Iraq مؤلف
  • Kadhim K. Resan Material Engineering Department, College of Engineering, Mustansiriyah University, Baghdad, Iraq مؤلف

DOI:

https://doi.org/10.31272/jeasd.26.4.7

الكلمات المفتاحية:

Amputee، shank، solid work، FGM، 2D printer، life shank device

الملخص

One of the most important problems that the components of the prosthetic limbs suffer from; is the increase in walking speed leads to an increase in the ground reaction force. In this research, a new shank was designed and manufactured using the porous functionally graded materials by a 3D printer of PLA material, and to ensure the success of this shank, several tests were conducted on it, including to know the mechanical properties, as well as the life test by a special device that was manufactured. A test was conducted to know the life of the shank when normal walking (about 1.35 m/sec) and running ( about 6 m / sec ), and it was found that this shank was successful and did not fail in that test, the prosthetic foot ( type SACH) failed before it at normal speed and high speed.                                              

التنزيلات

Key Dates

منشور

2022-07-01

كيفية الاقتباس

EFFECT OF THE GAIT SPEED ON A NEW PROSTHETIC SHANK BELOW KNEE. (2022). مجلة الهندسة والتنمية المستدامة, 26(4), 63-69. https://doi.org/10.31272/jeasd.26.4.7